# Neptune Auto Scan Settings Guide Source: https://support.lumafield.com/neptune/neptune-auto-scan-settings-guide Optimize your scans with Auto Scan ## What is Auto Scan? Auto Scan is Lumafield’s primary scanning method. Just put in your part, set the time, and let the system automatically optimize key scan settings to ensure the highest reconstruction quality, streamlining your workflow for routine inspections. ### Auto Scan vs. Manual Scan Lumafield recommends Auto Scan for the vast majority of applications. Most users will find that Auto Scan consistently delivers optimal results, with Manual Scan reserved only for highly specialized or non-standard inspection requirements. > [!NOTE] > > **Note:** For non-standard use cases that fall outside the scope of Auto Scan, please reach out to [support@lumafield.com](mailto:support@lumafield.com) for guidance on optimizing your workflow. ## Adjustable settings Auto Scan streamlines configuration so that specialized CT expertise is not required for high-quality imaging. Simply position your part, indicate the material composition, choose the scan duration, and let the Auto Tuner decide the rest of your scan parameters. ### Position and magnification ![Auto Scan Settings 1](https://lumafield-support-assets.s3.us-east-1.amazonaws.com/auto-scan-settings-1.png) Proper alignment is critical for resolution, as the Auto Tuner chooses settings based on what is in the Field of View (FOV). Use the interface controls to: * **Rotate:** Verify FOV clearance for a full 360° rotation * **Magnify (zoom):** Move the part closer to the source to increase resolution * **Translate (up/down):** Center the area of interest within the frame Ensure that components of the motion stage and turntable are outside the FOV for the best results. > [!NOTE] > > **Note:** Different Neptune models have specific "sweet spots" for magnification. For detailed resolution benchmarks, refer to the [**Neptune Positioning Guide**](https://support.lumafield.com/neptune/neptune-positioning-guide). This guide will also provide an overview of the advanced acquisition modules for Helical Scanning and Horizontal Offset Scanning. ### Adjust parameters ![Auto Scan Settings 2](https://lumafield-support-assets.s3.us-east-1.amazonaws.com/auto-scan-settings-2.png) **Part composition:** Mono-material versus Multi-material * **Mono-material:** Select this for parts with a single, uniform density. This setting is also appropriate for assemblies where the material densities are nearly identical, such as a component made from multiple similar plastics * **Multi-material:** Optimized for assemblies with a wide range of material densities, this setting helps the software better resolve high-contrast interfaces, such as metal components encased in plastic or complex PCBAs **Scan time:** Auto Scan durations can be configured for up to 12 hours. Increasing the scan time extends the total exposure, which reduces noise for superior reconstruction quality. While longer scans generally produce cleaner data, the impact varies based on a part's material and complexity. Note that the relationship between time and quality is not linear and that you are likely to encounter diminishing returns with increased time. As a general rule, the darker (more opaque) a part appears in the X-ray preview, the higher its density and X-ray attenuation. Higher attenuation parts require longer scan times to produce high quality reconstructions. For example, while homogenous plastic components may resolve in as little as 30 minutes, dense metal parts or complex assemblies often require longer scans (up to 12 hours) to ensure data clarity. ### Scan Sweep Scan Sweep is a process-optimization tool. It executes a series of back-to-back scans with varying durations within your total allotted Scan Time. Perform a Scan Sweep to compare quality across different scan times. * **One scan:** Standard single-acquisition mode * **Three or five scans:** Best for dialing in the minimum time required to achieve your desired image quality. All scans from the sweep will populate in the same project for ease of comparison across scan times ## Advanced Parameter Override ![Auto Scan Settings 4](https://lumafield-support-assets.s3.us-east-1.amazonaws.com/auto-scan-settings-4.png) The Advanced Parameter Override checkbox allows you to manually control some hardware settings that are typically managed by the Auto Scan algorithm. > [!NOTE] > > **Note:** These settings significantly impact scan physics. We recommend using these overrides only if you have a thorough understanding of CT acquisition or have consulted with the Lumafield Support team at [support@lumafield.com](mailto:support@lumafield.com). ### Filter selection Filters mitigate beam hardening artifacts by removing lower-energy X-rays before they reach the object. This ensures a more uniform beam that is better able to penetrate dense materials. * **When to use:** Helpful for multi-material assemblies and dense or metallic mono-material parts * **The trade-off:** Thicker filters require longer exposure times to maintain adequate signal, resulting in longer scan times * **Available options:** Copper filter thicknesses of 0.5 mm, 1 mm, 1.5 mm, 2.5 mm, and 6 mm ### Spot size (Neptune MFX only) Spot size determines the point of origin for the X-ray beam, creating a fundamental trade-off between geometric sharpness and power (flux). Auto Scan typically optimizes spot size to be close to the voxel size of the scan. * **Small Spot Size (5 µm):** Provides maximum resolution and sharpness; ideal for micro-features but requires longer scan times due to power limitations * **Large Spot Size (40 µm):** Allows for higher power and larger overall exposure. Suitable for larger, denser parts where micro-resolution is less critical. Can also be useful in increasing visibility of lower density plastics in a multi-material part * **Available Options:** 5 µm, 8 µm, 20 µm, and 40 µm # Scanning Electronics Quickstart Source: https://support.lumafield.com/neptune/neptune-electronics-guidance Practical Guidance for X-ray CT Scanning Electronics. X-ray CT is a powerful nondestructive inspection technique, but some electronics components can be sensitive to X-rays and other ionizing radiation.  Passive electronic components like resistors, capacitors, inductors, and the actual PCBs themselves are unaffected by exposure to ionizing radiation, but flash memory units, IMUs, and accelerometers are all susceptible to *total ionizing dose* (TID), which is a measure of the cumulative amount of ionizing radiation they have received. TID can cause electronic threshold shifts, leakage current, timing changes and functional failures. The total dose threshold varies by device, so it is best to determine the dose threshold for a given device via direct testing. When scanning electronics, it’s important to understand TID to perform a scan that both achieves a given inspection while minimizing the risk of affecting the device’s performance.   Dose is dependent on multiple factors, including: *  Scanner source energy *  Duration of the scan *  Number of times the device will be scanned *  Distance between the part and the source *  Amount of filtering used Dose rate is highest at high magnifications (smaller source to object distance) and at high source energy and lowest at low magnification (larger source to object distance) and low source energy.  The charts below show absorption rates for silicon, the primary material in active electronic devices, and can be used to determine the rate of absorbed dose at different magnifications with different source energies.    ![](https://lumafield-support-assets.s3.us-east-1.amazonaws.com/electronics-1.jpg) *Fig 1. Absorbed dose rate for electronics on 190 kV and 120 kV Neptunes.* ![](https://lumafield-support-assets.s3.us-east-1.amazonaws.com/Dose-Rate-For-MFX.png) *Fig 2. Absorbed dose rate for electronics on 130kV (MFX) Neptune.* The below table provides general guidance on TID thresholds for common electronic devices. | | | | - | - | | **Type of Semiconductor Device** | **Total Ionizing Dose Threshold (krads)** | | Linear | 2-50 | | Mixed Signal | 2-50 | | Flash Memory | 5-15 | | DRAM | 15-50 | | Microprocessors | 15-70 | *Fig. 2 X-ray Inspection Considerations for Surface Mounted Flash ICs. Infineon Technologies 001-98522 Rev. \*C [ref](https://www.infineon.com/dgdl/Infineon-AN98522_X-ray_Inspection_Considerations_for_Surface-Mounted_Flash_ICs-ApplicationNotes-v04_00-EN.pdf?fileId=8ac78c8c7cdc391c017d0742926a65a4).* ### Guidance for scanning electronics: * In practice, most scans of electronic devices will use some level of filtering to mitigate the artifacts that arise from multi-material assemblies.  Filtering reduces the amount of X-ray light that is used for a scan, reducing the dose, but longer scan times will typically be required when using filtering.  Refer to the charts above to understand the impact of the level of filtering on your scan. * While increasing magnification increases dose, it also increases fine feature resolution and reduces the chances of having to repeat a scan if the first scan does not resolve the target features. * Once you have positioned the part, set a scan duration using Auto Scan to find the other scan parameters.  Setting scan time in conjunction with the charts above will let you control the TID that the device receives.  ### Will my device be safe? A worked example As can be seen from the table above, TID thresholds for many devices vary dramatically.  While not always possible, the highest confidence method to assess the impact of X-rays on an electronic device is to find the TID limit experimentally.  See this worked example to learn how to determine the TID threshold experimentally.  **Device:** Flash Drive **Scan Conditions:** * Scanner Configuration: 190 kV * Source to Object Distance: 200 mm * Filter: 0.5 mm Copper * Scan Duration: 3 Hours **Procedure:** 1. Use the dose absorbance charts above to determine dose at the scan conditions.  In this example, the scan will be conducted on a 190 kV Neptune at a source to object distance of 200 mm with 0.5 mm of Copper filtering.  This corresponds to an absorbed dose rate of approximately 2 krad/hr.  A 3 hour scan would result in 3 hr \* 2 krad/hr = 6 krad of dose.  ![](https://lumafield-support-assets.s3.us-east-1.amazonaws.com/electronics-2.jpg) *Fig 3. Using the 190 kV absorption rate plot to identify the absorption dose rate at the intended scan settings.* 2. Scan the device in 1 krad increments and subject the device to functional testing in between doses.  In this example , this would mean scanning the device in 30 min increments. 3. Continue until the device fails functional testing. The TID at which the device fails testing is an experimentally-derived dose limit to which a margin of safety should be applied.  This margin can be itself experimentally determined by repeating the above procedure to build statistics about TID for the device.  ***As always, if you have any questions about your specific application, please contact us at [support@lumafield.com](mailto:support@lumafield.com)!*** # Expanded Scan Volume Quickstart Source: https://support.lumafield.com/neptune/neptune-expanded-volume How to use the offset motion axis to scan new objects. With the Expanded Scan Volume Module, we introduce new motion axis and scan modalities to the user.  Here is a video of how to use this feature and start scanning newer, larger objects. > *How to use the offset scan functionality.*