Skip to main content

Purpose of Application Development mode

Application Development mode allows you to perform custom CT scans outside of predetermined part programs. This flexible mode is ideal for research and development (R&D), inspecting single or one-off parts, and fine-tuning scan parameters to establish optimal settings for new product lines.

Access Requirement

Application Development mode is restricted to users logged in with Supervisor privileges. Once logged in, select Application Development at the bottom of the main menu.
Image

Step-by-step scanning procedure

  1. Fixture the part: Securely mount your part onto a conversion pallet
  2. Clear the stage: Press and hold the Air Release button to release the airlock on the rotary stage, then remove any existing pallet
  3. Load the pallet: Place your fixtured pallet onto the rotary stage
  4. Lock in place: Release the Air Release button to engage the airlock. Manually check the pallet to ensure it is fully locked and secure
  5. Configure settings: Select your desired scan parameters on the interface (refer to the Neptune Settings Guide for parameter recommendations)
  6. Start the scan: Review your parameters and initiate the scan